Signal-to-Noise
|
60000:1
1 min sample measurement, 4 cm-1 , peak-to-peak
|
Wavenumber precision
(@1554cm-1)
|
0.0005 cm-1
standard deviation of 10 repeated measurements
|
Spectral range
|
8,000 to 340 cm-1 standard KBr
optional ranges:
11,000 to 350 cm-1
5,000 to 200 cm-1
6,000 to 500 cm-1
|
8,000 to 350 cm-1 standard KBr
Optional extensions for far IR,
near IR and VIS/UV: maximum spectral range 28,000 to 15 cm-1
|
8,000 to 350 cm-1 standard KBr
Optional extensions for far IR,
near IR and VIS/UV: maximum spectral range 28,000 to 15 cm-1
|
Spectral resolution
|
Better than 0.4 cm-1
|
Better than 0.16 cm-1
|
Better than 0.085 cm-1
|
Wavenumber accuracy @1554cm-1
|
<0.01 cm-1 @ 1,554 cm-1
|
< 0.005 cm-1 @ 1,554 cm-1
|
< 0.005 cm-1 @ 1,554 cm-1
|
Temporal resolution @ 16 cm-1 spectral resolution
|
Up to 25 spectra per second
|
Up to 70 spectra/s with Rapid Scan option
|
Up to 70 spectra/s with Rapid Scan option
|
Temporal resolution with
Step Scan option
|
-
|
Up to 4 ns temporal resolution with
transient recorder board or 6 μs
using internal ADC.
|
Up to 4 ns temporal resolution with
transient recorder board or 6 μs
using internal ADC.
|
Interferometer
|
RockSolid™
|
RockSolid™
Beamsplitters are easy to exchange
and stored inside the optics bench
|
INTEGRAL™
equipped with 3-position automated
beamsplitter changer wheel
|
Detectors
|
DLaTGS-detector
Second detector position (option)
|
Up to 7 internal
1 external detector position
|
Up to 7 internal
1 external detector position
|
Light sources
|
MIR Globar
option for second NIR source
|
MIR Globar
option for second NIR/VIS source
option for external sources
|
MIR Globar
option for second NIR/VIS source
option for external sources
|
Upgrade options
|
Upgradable to INVENIO-R and INVENIO-X
|
Upgradable to INVENIO-X
|
-
|